LPS170-M Medical Series Artesyn 110-175 Watt Medical AC-DC Power Supplies
- по запросу
- Производитель: Artesyn
- Модель: LPS170-M Medical Series
Модель | Цена | На складе, шт | Количество |
LPS172-M | по запросу | по запросу |
|
LPS172-M-C | по запросу | по запросу |
|
LPS173-M | по запросу | по запросу |
|
LPS173-M-C | по запросу | по запросу |
|
LPS174-M | по запросу | по запросу |
|
LPS174-M-C | по запросу | по запросу |
|
LPS175-M | по запросу | по запросу |
|
LPS175-M-C | по запросу | по запросу |
|
LPS170-M series medically-approved AC-DC power supplies are available with a nominal main output of 5 V, 12 V, 15 V or 24 V, adjustable over a wide 2:1 ratio. Auxiliary 12 V and 5 V standby outputs are also provided. Main output remote sense and single-wire current sharing are supported as standard. LPS170 power supplies provide 110 watts of output power with free air convection cooling. which can be boosted to 175 watts with 30 CFM of forced air. All models have an 8.5 x 4.25 x 1.5 inch form factor and are ideal for low-profile space-constrained applications.
Особенности
- Medical safety approvals
- Built-in EMI filter
- 5 V standby output
- DC power good
- Optional cover (-C suffix)
- Thermal overload protection
- Overvoltage protection
- Low output ripple
- Active power factor correction
- IEC EN61000-3-2 compliance
- Wide Range Adjustable output
- Remote sense on main output
- 12 V Aux output
- Overload protection
- Power fail and remote inhibit
- Single wire current sharing
- Floating output
Технические характеристики
-
Total Max Power:
175 W -
Total Max Current:
22 A -
Input Voltage:
85 V to 264 V -
Output Voltage:
2.5 V to 54 V -
Outputs:
Single -
Size (L x W x H):
4.250" x 8.500" x 1.500" -
Warranty:
2 years -
Operating Temperature:
0 to +70°C
Electrical Specifications
- Input
- Output
- Safety ground leakage current: 275 uA @ 50/60 Hz, 264 VAC input
- EMI filter: Meets FCC Class A conducted CISPR 22 Class A
- Frequency: 47 - 63 Hz
- input range: 85 - 264 VAC, 120 - 300 VDC
- Inrush current: 38 A max, cold start @ 25°C
- Efficiency: 75% typical at full load
- Power factor: 0.99 typical
- Remote inhibit: Requires contact closure to inhibit outputs
- Aux output: 12 V @ 1 A -5 %, +10%
- Maximum power: 110 W convection, 75 W w/cover, 175 W w/ 30 CFM
- Overload protection: Short circuit protection on all outputs.
- Standby output: 5 V @ 2 A regulated ± 5%
- Hold-up time: 20 ms @175 W load at nominal line
- Overvoltage protection: 10% to 40% above nominal output
- Adjustment range: 2:1 wide ratio minimum
Список документов товара
- LPS172-M RoHS2 CoC
- LPS173-M RoHS2 CoC
- LPS174-M RoHS2 CoC
- LPS175-M RoHS2 CoC
- LPS174-M-C RoHS2 CoC
- LPS175-M-C RoHS2 CoC
- LPS170 CE Declaration of Conformity
- LPS170-M CE Declaration of Conformity
- LPS175-M UL/CSA ES60601-1 Medical Certificate ver 3.1
- LPS170-M TUV ITE/Medical Certificate
- LPS170-M Series Data Sheet